| Building a 4-Channel 2-Photon Microscope > Assembly and Fine-Tuning | |||||||||||||||||||
| Overview of Construction | |||||||||||||||||||
|
Getting
the electrical box built: Installing
the laser and routing optics and rough alignment: Paul Herzmark at UCBerkeley has compiled the following documents providing nifty and details from his own experiences aligning this path (click here). Installing
the shutter: |
|||||||||||||||||||
![]() |
|||||||||||||||||||
|
Note
that most of the items from within this dialog can be saved and retrieved
later using 'Save Settings' and 'Load Settings'. In addition, you will
see the VS main window: FYI, the tabs at the bottom are used to navigate
through the different functions of VS. |
|||||||||||||||||||
![]() |
|||||||||||||||||||
|
Configuring
the Camera File: |
|||||||||||||||||||
![]() |
|||||||||||||||||||
| Then
click on ‘Properties’. |
|||||||||||||||||||
![]() |
|||||||||||||||||||
| Choose
the appropriate camera file from the dropdown list : PMT_1500x200_4ch_HV_in.rvc
refers to the 4 PMTs (4-colors) system loaded into Bitflow using theSysReg
utility. You MUST also update the Vertical Image Size to a value equal to
1600. This sets the overall height of the initial images being captured. NOTE: You will get a message ‘Video Savant will now reset’ followed by ‘Cannot Register LUT Graph Class’ and you will be asked to verify the new dimensions. This is all normal—just click OK for each. 2.) Mounting your sample and getting samples oriented: A.) If necessary, drop a very small drop of water onto the sample and mount your sample in the appropriate sample carrier. (The 20x objective is a default and is water immersion—other lenses are available and see Max to use these). B.) With the eyepiece slider pulled out (output to eyes), turn on the transmitted light source for brightfield viewing. You can now use the focus knobs on the scope and the joystick for the stage to locate your sample. C.) Turn off the transmitted light and click ‘Open Shutter’ on the application page. You should now be able to focus through the eyepiece, using 2-photon illumination on the sample on which you focussed in B.). 3.) Using the PMTs. Switching to PMT Collection: In the confocal application, ‘close’ the shutter. Make sure the transmitted light source is ‘off’ and then move the slider so that emission goes to the PMTs. Failure to turn off the transmitted light source first can saturate the PMTs and you may have to wait minute/hours for them to calm down. Click the ‘Live’ Button. The shutter will open and images will start being collected at video rate. You should be able to view the image on the Video Savant screen. This ‘Live’ mode is useful for setting z-series and multiple stage positions (below). You can boost intensity by boosting PMT voltages (stay within the –400-->-1000 Volt range) or by integrating more frames. The Stream Filter and Displays: Behind the scenes, the video card is taking the input from each PMT and using a ‘stream filter’ in VS to convert the raster scan of the sample into images. This involves both interleaving the image as well as trimming data from the edge of the scan pattern where scan behavior is non-linear. The details of this stream filter can be seen by clicking ‘Stream Filter’ within the main Video Savant control panel. This bit of code is the brainchild of Mike Sanderson. Then click on the ‘Stream Filter’ on the left hand panel and click ‘Properties’. You should see the following dialog box. |
|||||||||||||||||||
![]() |
|||||||||||||||||||
| One
reason this scope is particularly quick is that data is collected on both
the forward and reverse paths of the scan mirrors. However, as the galvos
heat up on this scope and sometimes on a day to day basis, they may move
slightly faster or slower than predicted by the synchronization signals
that control them. A value of around 0 was initially set for the ‘Odd
Offset 1’ (in this screenshot, channel was set to offset much more.
This is highly atypical--usually these values are identical for all channels)
but values can change slightly from day to day. You’ll know this needs
to be changed since every other line will look a bit wobbly and some objects
will look like dual objects. A variety of channels can be displayed as a live upright file containing each color stacked vertically, a tiled set of up to four images, nd/or a small image containing a color composite. As opposed to the image in VS, this will always represent the output of the stream filtering process. Within the stream filter, you can choose which inputs (0-3) are displayed as which colors on a color-composite (using the dialogue) but only 3 colors max - use 0,1, 2 - R,G, B. The 4th would have to remain B/W in the alternate display modes.In addition, you can choose (see ‘setting up timelapses’ below) how many 1/30th second acquisition frames will be integrated to make the final image. This is ultimately the same number as the stream filter ‘average’ and does not need to be entered here unless you want to see the effects live (the acquire button places the value in the ‘Timelapse’ tab into this register—if you don’t hit acquire, you have to enter alternate averaging numbers in the box in the ‘stream filter’ dialog box. |
|||||||||||||||||||
|
Krummel
Lab • 513 Parnassus Avenue, Room HSW-518, San Francisco, CA 94143-0511
|
|||||||||||||||||||
|
Last Updated: September 19, 2007 • Webmaster: Christine W. Lin, christine.lin@ucsf.edu |
|||||||||||||||||||